EP Stimulator · The full simulator adds ICE, 3D conduction, fluoroscopy and more
See it conduct through the 3D heart → full simulator
Intervals in ms. Pause freezes the recording (Resume carries on, including any pacing); Stop pacing ends straight pacing; Pace & Pause pauses everything (like Pause) after the last extrastimulus. Choose the heart in Rhythm: the “(inducible …)” options start in sinus rhythm, ready for you to induce AVNRT, AVRT or focal AT; AVNRT, orthodromic and antidromic AVRT and typical flutter start already running (Cardiovert to stop them). For flutter, turn on the halo catheter, induce from CS proximal (burst ≤250 ms), entrain to read PPI − TCL, and use CTI ablation to test for bidirectional block; the atypical flutters (roof-dependent, clockwise and counter-clockwise peri-mitral) run or can be induced (burst from the distal CS), with a roof line or mitral isthmus line instead. VT (focal RVOT, scar re-entry) runs or can be induced: burst pacing ≤300 ms for focal VT, RV drive + S2 ≤270 ms for scar VT; overdrive pacing or Cardiovert terminates it. Atrial fibrillation starts running; Cardiovert, then burst atrial pacing at ≤200 ms re-induces it. Normal sinus rhythm has nothing to induce. HR-PVC: during a tachycardia, Deliver sends one RV extrastimulus timed to the expected His; the label then reports whether the next A was advanced, delayed, unchanged, or the tachycardia ended without one.